83
11-2 CALIBRATION RELATED FUNCTIONS
Setting
CALF- Name User’s
Default Parameter Description
0 None
1 g (International version)
2 kg ( International version and USA version)
3 t (International version)
01 Weighing Unit 2
4 lb ( USA version)
0 None 12345
1
10
1
1234.5
2
10
2
123.45
3
10
3
12.345
02
Decimal Point
Position
0
4
10
4
1.2345
03
Minimum
Division
1
1, 2, 5, 10,
20, or 50
Minimum division(increment) for the weight value.
Input 1,2, 5, 10, 20, or 50(decimal point ignored).
04 Capacity 16000 0 to 800000
Capacity of the weighing. Weighing is allowed up to
this setting + 9D(9 divisions). A weight value beyond
this limit is an overflow and not displayed.
05 Zero Range 2 0 to 30
A range of accepting "zero" from the [ZERO] key or
Control I/O. Represented in terms of percent(&) with
respect to the weighing capacity, centering around the
zero calibrated point in calibration. If this is set to 2,
"zero" can be accepted within a range of the zero
calibrated point +/- 2 %.
06 Zero Tracking Time 0.0 0.0 to 5.0
Zero tracking is performed in combination with
CALF-07 Zero Tracking Width. It is not performed
when set to 0.0. Its unit is seconds
07
Zero Tracking
Width
0 0 to 9
Zero tracking is performed in combination with
CALF-06 Zero Tracking Time. It is not performed
when set to 0. Its least input increment is 1/2D. (The
width for setting of 1 is equivalent to 1/2 of the
minimum graduation)
08 Motion Detect Time 1.0 0.0 to 5.0
Motion is detected in combination with CALF-09
Motion Detection Width. It is not detected when set to
0. Its unit is seconds.
09
Motion Detect
Range
2 0 to 9
Motion is detected in combination with CALF-08
Motion Detection Time. Its least input increment is 1D.
(The range for setting of 1 is equivalent to the
minimum division)
0
"Tare" is not accepted when the weight value is
unstable.
10*
Tare and Zero at
Unstable Weight
Value
1
1
"Tare" is accepted even if the weight
value is unstable.
0
"Tare" is not accepted when the
gross weight is negative.
11*
Tare at Negative
gross
1
1
"Tare" is accepted even if the gross
weight is negative.
Not available
on the USA
version.
Commentaires sur ces manuels