CHAPTER 10 MAINTENANCE
10-1 CHECK MODE
ENTERING CHECK MODE
The check mode is intended to check the operations of the indicator, key switches, and external input/output.
Step 1
With the [ENTER] key pressed and held, press the
[SETPOINT] key. “Fnc” is displayed to indicate that
the indicator is entering the general function mode.
Release both keys.
Note: When entering the function mode is not
wanted, press the [ESC] key. The indicator will
return to the normal mode.
Step 2
With the [ZERO] key pressed and held, press
the [ENTER] key. The indicator will enter the
check mode. “Chc key” is displayed and the
cursor appears at the right end. Release both
keys.
Step 3
Using the [ ] and [ ] keys, select the desired
check item.
Select the desired check item and press the
[ENTER] key. The indicator will enter the check
mode for the selected item. To exit each check
mode, press the [ESC] key.