27
Setting
FNCF- Name
Default Parameter Description
0 None
1 * discharging
05
Status display
Capability
0
2 Zero tracking
"Discharging" is exclusively intended for
normal batching (built-in automatic
program mode).
Value of each digit
and cutoff
frequency
0: None
1: 11.0Hz
2: 8.0Hz
3: 5.6Hz
4: 4.0Hz
5: 2.8Hz
6: 2.0Hz
7: 1.4Hz
8: 1.0Hz
06
Digital Filter
(Series
connection of
two digital
filters.)
48
00
to
79
9: 0.7Hz
The digital filter is designed to suppress
dispersion of a load cell output signal.
Two of them are connected in series as
shown below.
Low frequency components, which cannot
be covered only by setting of the digital
filters, can be dealt with by equivalently
lowering the cutoff frequency of the digital
filters at FNCF-07.
07
Sampling
frequency
dividing ratio.
1
1
to
10
Lower the cutoff frequency of the digital filters equivalently by
reducing the specified sampling times to one.
1 Normal hold
08 Hold operation 1
2 Peak hold
During peak hold, the positive peak of the
weight value is held.
0 No
09
Comparison
stop at “hold”
0
1 Yes
In Built-in automatic program mode, do not
set "1" because the weighing sequence
stops when comparison stops.