76
10-2 INITIALIZATION
Initialization is the operation to return the contents of the RAM and EEPROM to the initial factory values.
There are three kinds of initialization modes, depending upon their ranges.
RAM initialization mode
Initializes the RAM only. Since the zero compensation value, tare value, setpoint, accumulated
weight, and accumulated times are stored in the RAM, they are all cleared to zero.
General function initialization mode Initializes both the RAM and the general function settings stored in the EEPROM.
All data initialization mode
Initializes the RAM and all data stored in the EEPROM (the general function settings, calibration
related data) Since the calibration related data is also initialized, you must perform calibration
again.
RAM INITIALIZATION MODE
Step
1
Turn off the power. With the [SETPOINT] key pressed and held, turn on the power. All
display digits are illuminated, and then, “init ra” is displayed.
Step
2
Press the [ENTER] key. The indicator enters the RAM initialization mode. The “-” marks
appear above and below the upper and lower display sections to prompt you for
confirmation. When you want to initialize, continue pressing the [ENTER] key for 3
seconds or more. When you have entered this mode by mistake, do not hold the
[ENTER] key, press the [ESC] key to exit.
GENERAL FUNCTION INITIALIZATION MODE
Step
1
Turn off the power. With the [SETPOINT] key pressed and held, turn on the power. All
display digits are illuminated, and then, “init ra” is displayed.
Step
2
Press the [ ∧ ] key. “initFnc” is displayed.
Step
3
Press the [ENTER] key. The indicator enters the general function initialization mode. The
“-” marks appear above and below the upper and lower display sections to prompt you for
confirmation. When you want to initialize, continue pressing the [ENTER] key for 3
seconds or more. When you have entered this mode by mistake, do not hold the [ENTER]
key, press the [ESC] key to exit.
ALL DATA INITIALIZATION MODE
(Enter the All Data Initialization Mode from the calibration mode.)
Step
1
Remove the cover from the calibration switch located at the lower left of the front panel
and press the [CAL] key (found inside). “CAL” is displayed to inform you that the
indicator is entering the calibration mode.
Step
2
Press the [ENTER] key. Now, the indicator enters the calibration mode and “CAL SEt” is
displayed. The right most digit starts blinking.
Step
3
Press the [ ∨ ] key. “init ALL” is displayed
Step
4
Press the [ENTER] key. The indicator enters the all data initialization mode. The “-” marks
appear above and below the upper and lower display sections to prompt you for
confirmation. When you want to initialize, continue pressing the [ENTER] key for 3
seconds or more. When you have entered this mode by mistake, do not hold the
[ENTER] key, press the [ESC] key to exit.