Setting
CALF- Name
Default Parameter Description
08
Motion
Detection Time
1.0 0.0 to 5.0
Motion is detected in combination with CALF-09 Motion Detection
Width. It is not detected when set to 0. Its unit is seconds.
Motion
Detection
Range
2 0 to 9
Motion is detected in combination with CALF-08 Motion Detection
Time. Its least input increment is 1D. (The range for setting of 1
is equivalent to the minimum division)
09
0
"Tare" is not accepted when the weight
value is unstable.
10
Tare and Zero
at Unstable
Weight Value
1
1
"Tare" is accepted even if the weight
value is unstable.
0
"Tare" is not accepted when the gross
weight is negative.
11
Tare at
Negative gross
1
1
"Tare" is accepted even if the gross
weight is negative.
Not available on the
USA version.
0
1
Does not output when the weight value is overflowing or unstable
0
12
Standard Serial
Output; Output
When Weight
Value Is
Overflowing or
Unstable
1
1
Outputs even if the weight is overflowing or unstable
0 Does not output when the weight value is overflowing or unstable.
13
RS-232C/-422/-
485; Output
When Weight
Value Is
Overflowing or
Unstable
1
1 Outputs even if the weight value is overflowing or unstable
1
Normal batching
(Customer programmed control
mode)
2
Loss-in-weight
(Customer programmed control
mode)
3
Normal batching
(Built-in automatic program mode)
4
Loss-in-weight
(Built-in automatic program mode)
Batch
Weighing
For the
hopper
scale use
5 Check weighing 1
6 Check weighing 2
7 Check weighing 3
14 Weighing Mode 3
8 Check weighing 4
Check
Weighing
For the
platform
scale use