A&D OP-05 Spécifications Page 25

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 90
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 24
23
15
Zero Input
Voltage
0.000000
0.000000
to
2.200000
Input Voltage (mV/V) from the Load Cell at "Zero",
which is determined in "Zero Calibration" with weights.
16
Span Input
Voltage
(Capacity to
zero)
3.200000
0.000000
to
3.200000
Input Voltage (mV/V) from the Load Cell at "Span",
which means the difference between Capacity and
Zero. The voltage is determined in "Span Calibration"
with weights.
17
Weight against
Span Input
Voltage
16000
0 to 800000
(decimal point ignored)
When "Digital-Calibration", calibration not using
weights, is performed, CALF-15 and CALF-16
also have to be done.
This setting will be changed to Capacity (CAL-04) if
"CAL Set" is done.
Note: Write down the value of CALF-15, and
CALF-17 on the setting list attached back of this
manual for the maintenance purpose.
(Accuracy at replacement: approximately 1/500)
The scale can be calibrated with the stored
value, CAL-15,CALF-17. (Digital span function
approximately 1/1000) However it is
recommended to use weights for a better
accuracy.
Vue de la page 24
1 2 ... 20 21 22 23 24 25 26 27 28 29 30 ... 89 90

Commentaires sur ces manuels

Pas de commentaire