WEIGHING SEQUENCE FUNCTIONS
Setting
SQF- Name User’s Default
Parameter Description
1 Internal count
01
Selection of
comparison
weight
1
2 Display count
0 No Automatic accumulation
1 Accumulates only acceptable weight
02
Automatic
accumulation
0
2 Accumulates all values
0
No automatic free fall
compensation
1
Moving average of last
four times.
Only effective in Batch
weighing mode
03
Automatic free
fall
compensation
0
2
Fuzzy automatic free
fall compensation
Only effective in
Built-In-Automatic Program
mode.
04
Automatic free
fall effective
width.
0 0 to 99999999
Automatic free fall
compensation is made
if a loaded weight is
within the final +/-
automatic free fall
effective width.
Only effective in Batch
weighing mode.
05
Unstable
Dribble Flow
Timer
3.0
0.0 to 25.5
(In step of 0.1
second)
If Flow-time is shorter
than the setting, the
preset Free Fall is
applied to the weighing
instead of the
Auto-Free-Fall
compensation.
1 Real-time comparison
06
Overlimit/
underlimit
operation
2
2
Synchronized with
Batch Finish Output
0 Unrequired
07
Stability at
judgment
1
1 Required
0: Unused
08
Maximum
number of
supplementary
flow times
0 0 to255
Non-0: Set times
09
Batch start wait
timer
0.0
Only effective in
Built-In-Automatic Program
mode.
10
Full-flow
comparator
inhibitor timer
0.0
Only normal batching
(built-in automatic program
mode) is effective
11
Medium-flow
comparator
inhibitor timer
0.0
0.0 to 25.5
(In step of 0.1
second)
N/A
Only effective in
Built-In-Automatic Program
mode.