INSTRUCTION MANUAL 1WMPD4000506B AD-4406 OP-03/04/05/07/08/11 WEIGHING INDICATOR INSTRUCTION MANUAL
Page 8 AD-4406 Weighing Indicator 4. Description of Panels and Symbols 4.1. Front Panel Description No. Name Description 1 STABLE
AD-4406 Weighing Indicator Page 9 4.2. Rear Panel Description No. Name Description 19 Battery cover Battery compartment cover 20 Grou
Page 10 AD-4406 Weighing Indicator 5. Calibration The indicator converts an input voltage from a load cell to a "mass" value and display
AD-4406 Weighing Indicator Page 11 CAL Writes all the set data into non-volatile memory and displays CALoFF. Then, press the ON/OFF key to turn t
Page 12 AD-4406 Weighing Indicator Dual Range <First range> Specify the resolution, decimal point position and format. ↓ <First range>
AD-4406 Weighing Indicator Page 13 5.2.4. Specifying the Weighing Range of the First Range Step 3 After displaying CAP for 2 seconds, single ran
Page 14 AD-4406 Weighing Indicator F 2 ENTERF 2 ENTERStored input voltage New input voltage mV/V +/-ZEROTARE mV/V Place nothing on the weighing unit
AD-4406 Weighing Indicator Page 15 Weighing a Mass less than the Maximum Capacity Step12 Set a mass value using the following keys. key
Page 16 AD-4406 Weighing Indicator Press and hold the ON/OFF key and press the ESC key. No parameters are changed, CAnCEL is displayed
AD-4406 Weighing Indicator Page 17 5.3.1. Setting the Division and Range Consider the following rules to design the weighing range. Rule 1 Selec
This Manual and Warning Definitions The warnings described in this manual have the following meanings: CAUTION Disregarding the caution could
Page 18 AD-4406 Weighing Indicator 5.5. Gravity Compensation Function If the scale is used at the calibration location, it is not necessary to p
AD-4406 Weighing Indicator Page 19 5.5.1. The Gravity Acceleration Table Amsterdam 9.813 m/s2 Manila 9.784 m/s2 Athens 9.800 m/s2 Melbourne 9.8
Page 20 AD-4406 Weighing Indicator 5.6. Calibration Error Code List Exiting from a Calibration Error ESC key ... To return to the point where an
AD-4406 Weighing Indicator Page 21 6. Functions There are two parameter lists, one for the F-functions and one for CF-functions. These function
Page 22 AD-4406 Weighing Indicator 6.2. F-Functions Weighing Conditions (Digital filter, Zero tracking and Stability) Item Parameter Description 0
AD-4406 Weighing Indicator Page 23 Display and Other General Settings Item Parameter Description * 0 5 times/s F04 Display update rate 1 10 time
Page 24 AD-4406 Weighing Indicator External Input Item Parameter Description * 0 Not used (No function) 1 ZERO key 2 TARE key 3 NET/GROSS key 4 OFF
AD-4406 Weighing Indicator Page 25 Comparator Item Parameter Description * 0 Not used 1 Upper/lower limit comparison (2 Limits setting) 2 Upper/
Page 26 AD-4406 Weighing Indicator Data Output Item Parameter Description * 0 No data output 1 Analog output Set F31, 32, 33 2 Serial output F3
AD-4406 Weighing Indicator Page 27 Current Loop Output Item Parameter Description * 0 Displayed data 1 Gross data 2 Net data 3 Tare data F36 Out
AD-4406 Weighing Indicator Page 1 Contents 1. Compliance ...
Page 28 AD-4406 Weighing Indicator Serial Interface (continued) Item Parameter Description * 0 DP: point (.) / Delimiter: comma (,) F46 DP / Delimi
AD-4406 Weighing Indicator Page 29 6.3. CF-Functions Item Parameter Description * 0 No limitation CF00 Zero tracking width, motion detection cond
Page 30 AD-4406 Weighing Indicator 7. Function keys F1 and F2 For easier use, select functions of the F1 key and/or the F2 key and prioritize them
AD-4406 Weighing Indicator Page 31 8. Tare The function is used to display a net value with the container weight subtracted from the total weight,
Page 32 AD-4406 Weighing Indicator 9. Accumulation The function accumulates weight data and stores the total data and the accumulation count. Dat
AD-4406 Weighing Indicator Page 33 9.2. Display and Operation Action of Accumulating Data When accumulating data, the display blinks once.
Page 34 AD-4406 Weighing Indicator Output of Accumulated Data The accumulated data can be output to the serial interface. Output by manual or
AD-4406 Weighing Indicator Page 35 10. Code Memory The AD-4406 has four Code memories (1 through 4). Each Code memory stores a set of setpoints.
Page 36 AD-4406 Weighing Indicator 11. Comparison This function has the "upper / lower limit comparison", the "5-stage (HiHi/Hi/O
AD-4406 Weighing Indicator Page 37 11.1.1. Condition Formula for Comparison Comparison is performed based on the following formula. Upper/lower lim
Page 2 AD-4406 Weighing Indicator 6.1. Changing the Function Settings ...
Page 38 AD-4406 Weighing Indicator Step 3 Select the comparison class (upper limit value, etc.) using the following keys. key ... To select
AD-4406 Weighing Indicator Page 39 11.2. Setpoint Comparison This function includes the weighing sequence and uses for acquiring a preset targe
Page 40 AD-4406 Weighing Indicator F22 = 9 (Check weighing 3: Loss in weight) Output and Condition * Relay OK output can be changed
AD-4406 Weighing Indicator Page 41 Toward the Zero band Preliminary and Free fall output are holding the off state. Over/Under comparison st
Page 42 AD-4406 Weighing Indicator 12. Hold Function This function displays the hold weight data after averaging the weight data for a specific
AD-4406 Weighing Indicator Page 43 The key switch will function as the HOLD key if F10 or F11 parameter is set to A. The external input func
Page 44 AD-4406 Weighing Indicator 13. RS-232C Interface (OP-04, OP-05, OP-08) 13.1. Specifications Transmission Asynchronous, bi-directional, half
AD-4406 Weighing Indicator Page 45 13.2. Data Format There are two types of data format set at F-Function F34 and F35. The initial data form
Page 46 AD-4406 Weighing Indicator When receiving an undefined command (unknown command), a "?" is sent back. When the memory is sh
AD-4406 Weighing Indicator Page 47 Request Net Data When receiving this command, returns the net data immediately. Template RN Command Rep
AD-4406 Weighing Indicator Page 3 1. Compliance 1.1. Compliance with FCC Rules Please note that this equipment generates, uses and can radiate r
Page 48 AD-4406 Weighing Indicator Display Gross Data Displays the gross data. Template MG Command Reply Display Net Data Displays the
AD-4406 Weighing Indicator Page 49 Enabling Key Switches Enables the key switches that are disabled by the DK command. Not applicable to the key
Page 50 AD-4406 Weighing Indicator Request Hold State When receiving this command, returns the average/hold state immediately. Template HD Co
AD-4406 Weighing Indicator Page 51 data Weight, result of comparison etc. $CL CLear previous settings. The UFC command parameters can not be cha
Page 52 AD-4406 Weighing Indicator 14. RS-422/RS-485, Relay Output (OP-03) Replacing the RS-232C interface with this option, the RS-422/RS-485 int
AD-4406 Weighing Indicator Page 53 Switching Between RS-422/RS-485 Switching between RS-422/RS-485 is made with the slide switch (S1) on the OP-03
Page 54 AD-4406 Weighing Indicator 15. Relay Output & Control Input (OP-05) Replacing the RS-232C interface with this option, 3-relay outputs
AD-4406 Weighing Indicator Page 55 16. 4-20 mA Analog Output (OP-07) The OP-07 analog output option is for sending the weight data to an analog
Page 56 AD-4406 Weighing Indicator 17. Current Loop Output (OP-08) Replacing the RS-232C interface with this option, current loop output, 3-relay o
AD-4406 Weighing Indicator Page 57 Current Loop Output The current loop output can be used to output data to an A&D printer and a display uni
Page 4 AD-4406 Weighing Indicator 2. Introduction The AD-4406 is a compact weighing indicator that amplifies the signal from a load cell, convert
Page 58 AD-4406 Weighing Indicator 18. Specifications Analog Input and A/D Conversion Input sensitivity Up to 0.2 µV/division Input signal range -
AD-4406 Weighing Indicator Page 59 18.1. Dimensions 18.1. Dimensions
AD-4406 Weighing Indicator Page 5 3. Installation and Precautions 3.1. Installation and Precautions The AD-4406 weighing indicator is a precisi
Page 6 AD-4406 Weighing Indicator 3.3. The Load Cell Connections Connect the load cell wires to the connector (receptacle), at the rear panel, usi
AD-4406 Weighing Indicator Page 7 3.5. Verifying Load Cell Output and Input Sensitivity The input sensitivity of the indicator is 0.2 µV/division