A&D OP-05 Spécifications Page 64

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 90
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 63
62
DATA TRANSFER MODE
There are three kinds of data transfer modes for the BCD output; “stream,” “auto print,” and “manual print.”
[Stream], [Auto Print], and [Manual Print]
The functions of these transfer modes are the same as described in {8-3 Standard Serial Output}.
TIMING
Strobe pulses are turned on approximately 7 ms after rewriting the output data and turned off in 17 ms.
If a hold input is turned on , an update of the output data will stop within 50 ms.
9-2 OP-03 RS-422/-485 INTERFACE
OP-03 provides external inputs/outputs to be connected to a personal computer. It can be set to RS-422 or RS-485. It
is different from the RS-232C in that up to 10 units can be connected in series (RS-422) or parallel (RS-485) by the
address capability. The data format, function settings, and so on are common to the OP-04 RS-232C.
Signal
System
Conforms to the EIA RS-422 or RS-485
Data bits 7 bits, 8 bits
Parity bits 1 bit (Even or Odd), None
Stop bits 1 bit, 2 bits
Baud rate 600, 1200, 2400, 4800, 9600, 19200 bps
Code used ASCII
ADDRESS CAPABILITY
The address capability responds only when your own
instrument is called from a host computer. The address
number of each instrument can be set from 1 to 99 with
RSF-09; 0 denotes no address capability.
When a command prefixed with an address “@xx”(xx is an
address number) is sent from the host computer, each AD-4401 compares this with address setting of the individual
instrument.
If a result of setpoint matches, the command will be analyzed and responded to. At this time, the instrument’s address is
added to the response.
Example of Adding the Address (For MZ Command)
SWITCHING TO RS-422/RS-485
Switching to RS-422/RS-485 is made with RSF-08. When more than two units are connected to the AD-4401, select "
Command " for RSF-02 and " With address " for RSF-09.
Vue de la page 63
1 2 ... 59 60 61 62 63 64 65 66 67 68 69 ... 89 90

Commentaires sur ces manuels

Pas de commentaire