Entering the density directly (Ld in 1)
In the specific gravity measuring mode, press and hold the [MODE] key until the density currently set
(unit: g/cm
3
, factory setting: 1,0000g/cm
3
is displayed. Use the following keys to change the value.
[RE-ZERO] key Changes the numerical value of the digit selected.
[RANGE] key Selects the digit to change the value.
[PRINT] key Saves the change, displays „End“ and returns to the specific gravity
measuring mode
[CAL] key Returns to the specific gravity measuring mode without saving the change.
Note
• The range to set the density is 0,0000-1,9999/cm
3.
(Displayed up to four decimal places)
Measuring the density
In the density measurement, the balance displays the weight of the sample in air, the weight in liquid
and then the density.
• Measuring the weight of the sample in air.
The processing indicator flashes with the unit „g“ displayed.
• Measuring the weight of the sample in liquid.
The processing indicator illuminates with the unit „g“ displayed.
• Displaying the density.
The processing indicator illuminates with no unit displayed.
To switch between the above three, use the [RANGE] key.