English
GB 39
Interval Memory Mode
This is the data memory function mode. Weighing data is periodically stored in the balance. The interval
memory mode can not be used, while stream mode is used.
Required setting dout Prt 3 Stream mode
dout dAtA 1 Data memory function is used
dout int Interval time
Example “Periodical weighing without computer command and outputting all of the data
to a computer at one time.”
9.2. “Data format”
A&D standard format SiF tyPE 0
This format is used when the peripheral equipment, capable of receiving A&D format, is connected. If an
AD-8121 is used, set the printer to mode 1 or 2.
• This format consists of fifteen characters (excluding the terminator).
• A header of two characters indicates the status of the stability.
• The plus sign is placed before the data, when the data is zero or positive.
• The weight data uses leading zeros.
• The unit has three characters.
Header
Data Unit
Terminator
Stable header
Unstable header
Overload header
Stable header for counting mode
D.P. (Dump print) format SiF tyPE 1
This format is used when the peripheral equipment can not process the A&D format. If an AD-8121 is
used, set the printer to mode 3.
• This format consists of sixteen characters (excluding the terminator).
• A header of two characters indicates the status of the stability without overload.
• The polarity sign is placed before data, is not zero or overloaded.
• The weight data has spaces in place of the leading zeros.
• The unit has three characters.
Header
Data Unit
Terminator
Stable header
Unstable header
Stable header for counting mode
Commentaires sur ces manuels