3.5. Calibration Using an External Weight
English
GB 19
Operation
1. Connect the AC adapter and warm up the balance for at least
one hour with nothing on the pan.
2. Press and hold the [CAL] key until displaying CAL out and
then release the key.
3. The balance displays CAL 0.
If you want to change the calibration weight value,
proceed to step 4.
If you use the stored calibration weight value in the
balance, proceed to step 5.
4. Press the [RANGE] key and adjust the calibration weight
using the following keys.
[RE-ZERO] key The key to set the value of the digit selected.
[RANGE] key The key to select the digit to change value.
[PRINT] key The key to store a new weight value and
return to step 3.
[CAL] key The key to cancel this change and return to
step 3.
Product Usable external weight Adjustable range
G2105D
200 g, 100g
G2104 +15.9 mg ~ –15.0 mg
G1204 100 g, 50 g
GR-300 200 g, 300 g
Digits cyclically change using the [RE-ZERO] key.
ex. 0 mg › +15 mg › –15 mg › 0 mg
5. Confirm that there is nothing on the pan and press the [PRINT]
key. The balance measures the zero-point. Prevent vibration and
draughts from affecting the balance.
Press and hold the key
Release the key
ex. calibration weight
value = 200.0017 g
Initial calibration
weight value is
200.0000 g
at GR-200.
To Next Page