A&D GH Series Manuel d'instructions Page 45

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 71
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 44
10.4. Calibration test report using the internal weight
English
GB 45
Key operation
1. Press and hold the [CAL] key until displaying CC in. Release the key.
2. The balance displays CC and performs calibration test automatically.
3. The zero point is measured and this value is displayed.
4. The internal weight is measured and this value is displayed.
5. If the calibration test report is output, GLP is displayed and the GLP data is output.
6. The balance returns to the weighing mode automatically.
Command operation
1. Transmit the TST command to the balance.
2. The balance performs the calibration test automatically.
3. If the calibration test report is output, the GLP data is output.
4. The balance returns to the weighing mode automatically.
Vue de la page 44
1 2 ... 40 41 42 43 44 45 46 47 48 49 50 ... 70 71

Commentaires sur ces manuels

Pas de commentaire