A&D GH Series Manuel d'instructions Page 24

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 71
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 23
A&D GR Series Instruction Manual
GB 24
Calibration
From Previous Page
8. Press the [PRINT] key to enter into the procedure for correcting the
internal weight value.
9. Correct the internal weight value using the following keys.
[RE-ZERO] key The value is selected.
(+1.5 mg ~ –1.5 mg)
[PRINT] key The new value is stored and iD is displayed.
[CAL] key This correction is cancelled and iD is displayed.
10.Press the [CAL] key. The balance will return to the weighing mode.
11.Press the [CAL] key to calibrate the balance.
Control of the Correction
Correction of the internal weight value is controlled by the parameters
“Permission or Prohibition”. Refer to section “7. Function Switch and
Initialization”.
Function table
0 : Change not permitted
1 : Change permitted
Correction of the internal weight
0 : Not used
1 : Used
Vue de la page 23
1 2 ... 19 20 21 22 23 24 25 26 27 28 29 ... 70 71

Commentaires sur ces manuels

Pas de commentaire