A&D GR Series Instruction Manual
GB 44
10.2. GLP output
10.3. Calibration report using the internal weight
Set the following parameters to output the report:
If the report is printed, set the “GLP output (inFo)” to “1”. The AD-8121 printer is used in this
explanation. Refer to “14. Connection to the AD-8121”. The AD-8121 uses MODE 3.
If the report is output to the RS-232C interface of a computer, set the “GLP output (inFo)” to “2”.
Key operation
1. Press the [CAL] key to display cAL in. The balance calibrates automatically.
2. If the calibration report is output, GLP is displayed and the GLP data is output.
3. The balance returns to the normal weighing mode automatically.