A&D GH Series Manuel d'instructions Page 20

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 71
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 19
A&D GR Series Instruction Manual
GB 20
6. Place the displayed calibration weight on the pan and press the
[PRINT] key. The balance displays the measured calibration
weight.
7. Remove the weight from the pan after the balance displays
End.
8. If the “GLP output (inFo)” of the “Function Table” is set to “/
or“2”, “Calibration Test Report” is output by the RS-232C
interface.
9. The balance will automatically return to the weighing mode after
calibration.
10. Test the accuracy of weighing using the calibration test function
with a certified test weight.
Controls of this Calibration
Calibration using an external weight is controlled by a parameter
in “Permission or Prohibition” Refer to Section “7. Function
Switch and Initialization”.
From Previous Page
Calibration Using the External Mass
0 : Not used
1 : Used
Calibration
weight
GLP output
Calibration report
Vue de la page 19
1 2 ... 15 16 17 18 19 20 21 22 23 24 25 ... 70 71

Commentaires sur ces manuels

Pas de commentaire