A&D GR Series Instruction Manual
GB 22
6. Place the displayed target weight on the pan and press the
[PRINT] key. The balance displays the measured target weight
and displays it. Prevent vibration and draughts from affecting the
balance.
7. Remove the weight from the pan after the balance displays End.
8. If the “GLP output (inFo)” of the “Function Table” is set to “/” or“2”,
“Calibration Test Report” is output by the RS-232C interface.
9. The balance will automatically return to the weighing mode after
calibration.
From Last Page
Target
weight
GLP output
Calibration test report