AD-4406 Weighing Indicator Page 39
11.2. Setpoint Comparison
This function includes the weighing sequence and uses for acquiring a preset
target weight.
There are four parameters of "Final", "Preliminary", "Free fall" and "Zero band" to
use the setpoint comparison.
The result of the sequence is output to three relays of OP-03, OP-05 or OP-08.
When entering these parameters, it is not necessary to enter the F-function F22
again unless comparison conditions are changed.
11.2.1. Description of Input Parameters and Outputs
F22 = 7 (Check weighing 1)
Output and Condition
Relay OK output can be changed to OVER or UNDER by setting F24. The
triangle 2 is not displayed.
OVER/UNDER is judged always and the result is output to the display and the relay.
F22 = 8 (Check weighing 2)
Output and Condition
OVER/UNDER is judged always and the result is output to the display but not to
the relay.
11.2. Setpoint Comparison
Final
Weighing value
Calibrated Zero
Tare
Tared Zero
Zero band
Free fall
Preset target weight
Time
Weighing trace
Relay HI, 1, Net (Final - Free fall)
Relay LO, 3, Gross Zero band
ON
ON
ON
OFF
OFF
OFF
Final
Preliminar
y
Weighing value
Calibrated Zero
Tare
Tared Zero
Zero band
Free fall
Preset target weight
Time
Weighing trace
Relay HI, 1, Net (Final - Free fall)
Relay OK, 2, Net (Final - Preliminary)
Relay LO, 3, Gross Zero band
ON
ON
ON
ON
OFF
OFF
OFF
OFF
OFF