AD-4406 Weighing Indicator Page 29
6.3. CF-Functions
Item
Parameter
Description
* 0 No limitation
CF00
Zero tracking width,
motion detection
condition
1 Use limitation at F01, F02, F03, F27 and F28
* 0 2% of CAP, Tare limit is 100 % CAP
1 10% of CAP, Tare limit is 100 % CAP
2 3% of CAP, Tare limit is 50 % CAP
CF01
Push zero range
3 4% of CAP, Tare limit is 50 % CAP
0 Not to zero when turning the display on.
*1 10% of CAP
2 3% of CAP
CF02
Power on zero range
Turning the display on,
the range to zero display.
3 4% of CAP
0 Gross when displaying gross
1 Gross
CF03
Zero tracking
* 2 Gross or Net when displaying net
TARE, ZERO in motion / TARE at negative gross
* 0 Not to execute / Not to execute
1 Execute / Not to execute
2 Not to execute / Execute
CF04
TARE, ZERO in motion /
TARE at negative gross
value
3 Execute / Execute
* 0
Not to output data at unstable value or overload
Effective in key mode
CF05
Output on overload and
unstable state.
1 To output data always
GROSS / NET / TARE
* 0 GS / NT / TR
1 GS / NT / TR
CF06
Header 2
2 G_ / N_ / T_ (_:Space 20h)
* 0 Two digits CF07
Number of unit digits
1 Three digits
* 0 Not used (Ineffective) CF08
Accumulation function
1 Used (Effective)
CAP : maximum capacity *: Initial settings
Power on ZERO (CF02)
CF02=0 After power on, the weight display starts immediately.
CF02=1,2,3 When the zero range is exceeded at power on, ------ is displayed.
Press the ESC key to start the weight display.
6.3. CF-Functions