AD-4430B - 41 -
7.5. Verifying The Load cell Connections Using Multimeter
The load cell connection can be checked easily using a digital multimeter.
Illustration 12 Connection Check of Load Cell shows points to confirm the load cell connection.
When a summing box is used, the same measurement must be performed, even internally.
Illustration 12 Connection Check of Load Cell
7.5.1. Check List Of The Load cell Connections
Measurement
Point
Item To Be Measured Judging The Voltage
EXC+ SEN+
A drop voltage of the load
cell cable on EXC+ side.
Normally 100mV or less. However, it may exceed 1V
when an extremely long load cell cable is used. For the
4-wire connection, it must be 0V.
EXC+ EXC- Input voltage of load cell Normal range is between 4.75 V to 5.25 V.
SEN- EXC-
A drop voltage of the load
cell cable on EXC- side.
Normally 100mV or less. However, it may exceed 1V
when an extremely long load cell cable is used. For the
4-wire connection, it must be 0V.
SIG- EXC- Center point voltage Approximately 2.5 V of a half of excitation voltage.
SIG+ SIG- Output voltage of load cell
Generally within 0 V to 15mV. Theoretical value is
calculated from the load cell rated capacity, actual load
and excitation voltage.
O
utput voltage of load cell
Center point voltage
Drop voltage between
EXC- and SEN-.
Input voltage of load cell
Drop voltage between
EXC+ and SEN+.
Load cell
Shield cable
SIG-
EXC-
SEN-
SIG
+
SEN
+
EXC
+