12. SPECIFICATIONS
AD-4212B-PT AD-4212A-PT
FX-300i-PT
Weighing capacity
*1
110 g / 31 g / 5.1 g
*2
110 g 320 g
Minimum weighing
value
0.1 mg / 0.01 mg / 0.001 mg 0.1 mg 1 mg
Linearity
±0.2 mg / ±0.05 mg / ±0.05 mg ±0.3 mg ±2 mg
Repeatability
(Standard deviation)
0.1 mg / 0.05 mg / 0.015 mg 0.15 mg 1 mg
Dimensions
Weighing unit: 80 (W) x 230 (D) x 200 (H) mm
Display (with a stand): 237 (W) x 150 (D) x 155 (H) mm
193 (W) x 262.5 (D)
x190 (H) mm
Standard
accessories *3
Instruction manual
Balance including the weighing pan unit, breeze break, AC adapter and AC
adapter ID label
Calibration weight with a pair of tweezers
Evaporation trap
Sample cup (30 mL x 2 / 5 mL x 2)
*4
Sample cup holder (30 mL / 5 mL)
Thermometer
USB communications kit (USB converter, RS-232C cable, Instruction manual)
WinCT-Pipette (CD-ROM)
Carrying case with a shoulder belt and a key
AC adapter Please confirm that the AC adapter type is correct for your local voltage and
power receptacle type.
Pr consumption Approx. 11VA (supplied to the AC adapter)
Carrying case
dimensions
470 (W) x 150 (D) x 355 (H) mm
Mass
(With
all accessories
in a case)
Approx. 7.6 kg Approx. 7.2 kg Approx. 6.4 kg
*1 When the balance weighing pan is used.
*2 The AD-4212B-PT is equipped with a smart range function. When the minimum weighing value is
set to 0.001 mg and the mass value exceeds 5.1 g, the minimum weighing value will switch to
0.01 mg, and when the mass value exceeds 31 g, the minimum weighing value will switch to 0.1
mg automatically. Even under this circumstance, pressing the RE-ZERO key tares the value and
weighing with the minimum weighing value of 0.001 mg is available from that point up to 5.1 g.
*3 The standard accessories for the AD-4212B-PT / AD-4212A-PT / FX-300i-PT are different from
those for the AD-4212B / AD-4212A / FX-300i.
*4 Sample cups and water-absorbent sheets are available separately for purchase as options.