11. TROUBLESHOOTING
Displayed values are unstable.
The AD-4212B-PT / AD-4212A-PT has a minimum weighing value of 0.001 mg / 0.1 mg and is
influenced by the measurement environment.
Measures to take
Use the accessory breeze break.
Referring to “3. PRECAUTIONS BEFORE USE”, check the measurement environment for
sources of drafts, vibration and static electricity.
The AD-4212B-PT with a minimum weighing value of 0.001 mg is prone to low-frequency
vibration such as shaking buildings. Particularly, earthquake, wind and changes in the
barometric pressure influence measurement. To avoid vibration, install the balance on the first
floor, in the corners of rooms (close to pillars or walls) far from a passage way.
Use the AD-1685 anti-vibration table.
Use the optional AD-1683 DC static eliminator or use the optional AD-1684 electrostatic field
meter if the cause of the unstable values may be due to static charges.
Repeatability of the measurement values is low. The measurement values are not correct.
Measures to take
Referring to “3. PRECAUTIONS BEFORE USE”, check the measurement environment. Using
the accessory calibration weight, measure the balance repeatability and check the balance
performance.
Referring to “9-3 Example Causes of Measurement Errors”, check the measurement
environment for causes of errors.
Balance error displays
Refer to the balance instruction manual.