8.9.
8.9.8.9.
8.9. Analog Output of Option OP
Analog Output of Option OPAnalog Output of Option OP
Analog Output of Option OP-
--
-07
0707
07
This option outputs DC current that is proportion to the display value.
Factory adjusted to 4 mA output at zero display and 20 mA output at full scale.
Analog output Contact to ground
Output voltage 11 V DC min.
Adaptable resistance 0 ~ 500
Update ratio 100 times per second with Sampling frequncy divider
[genf- 3]
Zero temperature coefficient 150 ppm/ max.
Span temperature coefficient 150 ppm/ max.
Non-linearlity 0.1% max.
Resolution Smaller value of either 1/40000 or resolution of display
Connection
Settings of Parameters
Refer to "10.4. Parameter List" of the function list.
Span adjustment screw
adjusted in factory.
Zero adjustment screw
adjusted in factory.
AD-4402
OP-07 inside
2
Constant current source
10V max.
FG
3
FG
1
The output voltage varies
according to resistor value.
Output current
Resistor
+
-
Connect shield
100.00kg
When setting to
[
07 f- 2] [0.00]
[
07 f- 3] [100.00]
0.00kg
20mA
4mA
2mA
22mA
When setting to
[
07 f- 2] [100.00]
[
07 f- 3] [0.00]
Weighing
Output current
50.00kg
When setting to
[
07 f- 2] [0.00]
[
07 f- 3] [50.00]