A&D Multi Function Weighing Indicator AD-4402 Spécifications Page 96

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 146
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 95
Page 94 AD-4402
8.8.
8.8.8.8.
8.8. Parallel I/O of Option OP
Parallel I/O of Option OPParallel I/O of Option OP
Parallel I/O of Option OP-
--
-05
0505
05
Use this option to extend the I/O terminals
The function, settings, interface circuit and timing chart of the option is the same as
the built-in I/O terminal.
Input control Contact to common
Input open voltage 7 ~ 11 V DC
Input current 5 mA max.
Input threshold voltage 2 V max.
Output circuit Open collector transistor
Output voltage 40 V DC max.
Output saturation voltage 1.5 V at 50 mA
Connection
A1 ~ A16 Input terminals
A17
A18
A19
Input common
A20 Frame ground
B1 ~ B16 Output terminals
B17
B18
B19
Output common
B20 Frame ground
Terminal List
Refer to "10.4. Parameter List" of the function list.
Caution
Do not assign the same function to multiple input terminals and keys.