4-2. Setting a weighing instrument and OP-08
(1) Setting a weighing instrument
Set the internal setting of the serial interface of the weighing instrument as follows:
This setting is the same as the factory setting.
Setting item Setting value
bPS
Baud rate
2400 bps
btPr
Data bit , parity bit
7 bit EVEN
(2) Setting an OP-08
Use “DeviceInstaller” software for setting.
Cancel the firewall function on the computer.
Step 1 Connect the computer and the weighing instrument with the OP-08 to the same network.
If multiple instruments are connected, connect one by one.
Factory setting is set at the same IP address (172.16.100.2).
Step 2 Start from the Start menu on the computer.
Click “Start”, “Program”, “Lantronix”, “DeviceInstaller” to start.
Step 3 Click “Search” to perform automatic detection of the OP-08 connected to the network.
Step 4 Click on the (+) mark to expand the parameters and select the IP address that you would
like to change as shown in the diagram below. The factory default number is 172.16.100.2.