28
AD-8121 format “info 1 General format “info 2
: Space, ASCII 20h
CR: Carriage return, ASCII 0Dh
LF: Line feed, ASCII 0Dh
Output of “Title block” and “End block”
When a weight value is recorded as the GLP report, “Title block” and “End block” are
added at the beginning and at the end of a group of weight values.
Title block
1. Press and hold the PRINT key. Release the
PRINT key when start is displayed. The
balance outputs the “Title block”.
2. The balance can output the weight data by
pressing the PRINT key or selecting the auto-
print mode.
A & D
MODEL EK-1200i
S/N Q1234567
ID ABCDEF
DATE 2011/11/14
TIME 03:15:40
CAL.TEST(EXT.)
ACTUAL
0.0 g
+1000.1 g
TARGET
+1000.0 g
SIGNATURE
- - - - - - - -
~~~~~~~~~~~A~&~D<CRLF>
MODEL~~~EK-1200i<CRLF>
S/N~~~~~Q1234567<CRLF>
ID~~~~~~~~ABCDEF<CRLF>
DATE<CRLF>
<CRLF>
TIME<CRLF>
<CRLF>
CAL.TEST(EXT.)<CRLF>
ACTUAL<CRLF>
~~~~~~~~~~0.0~~g<CRLF>
~~~~~~+1000.1~~g<CRLF>
TARGET<CRLF>
~~~~~~+1000.0~~g<CRLF>
SIGNATURE<CRLF>
<CRLF>
<CRLF>
---------------~<CRLF>
<CRLF>
<CRLF>
end
start
0.0
To output
“Title block”
123.1
123.4
To output
weight data
Press and hold
Manufacturer
Model
Serial numbe
r
ID number
Date
Time
Calibration
test
Target weight value
Column for
signature
Zero value
Actual weight
value