AD-4408C Weighing Indicator Page 58
11. Setting List
When performing maintenance, use the following list as a memo.
In case of inquiry about the product, inform of the user setting.
11.1. Basic Functions
Function No.
Parameter or
Setting range
Description
Default
value
User
setting
Ffncf01
000000 to
111111
Disable key switch
0: Does not disable the key switch function.
1: Disables the key switch function.
000000
(binary)
fncf02
0 to 6
Assigns a function to the F key.
0: None
1: Manual print command
2: Hold
3: Alternate switch
4: Momentary switch
5: Clear the tare weight
6: Clear the zero value
0
fncf03
1 to 3
Display update rate
1: 20 times/second
2: 10 times/second
3: 5 times/second
1
fncf04
0 to 3
Assigns a function to the display.
0: None
1: Zero tracking in progress
2: Alarm (Zero range setting error, over,
low battery)
3: F key status
0
fncf05
0 to 9
Digital filter (cutoff frequency)
0: None
1: 11.0 Hz
2: 8.0 Hz
3: 5.6 Hz
4: 4.0 Hz
5: 2.8 Hz
6: 2.0 Hz
7: 1.4 Hz
8: 1.0 Hz
9: 0.7 Hz
8
fncf06
1 to 10
Frequency divider ratio
1
fncf07
1 to 2
Hold
1: Normal hold
2: Peak hold
1