ScaleResource.com 1A&D 4321 1. With indicator in standby mode 2. Remove plate on front of indicator 3. Flip cal switch up 4. Push standby/o
ScaleResource.com 104. Press enter then enter a 0 to bypass or 1 to enter cal. 5. Press enter, display shows C1, press enter, a number is display
ScaleResource.com 119. St. 0 0 10. Press on/off key. CAS S2000 1. Turn scale off, remove platter. 2. Remove the cover from the cal access port
ScaleResource.com 12 CAS AP-1 1. Remove cal bolt on rear of scale, flip cal switch to left. 2. Put platter on scale. 3. Apply power, Flashes CA
ScaleResource.com 137 SEQUENCE # 0001-9999 8 STATION # 00
ScaleResource.com 1410. TO GO DOWN, PUT THE LB/KG SWITCH IN KG MODE AND CLICK THE ON/ZERO-OFF SWITCH TO “ZERO”. 11. WHEN THE DISPLAYED WEIGHT MATCH
ScaleResource.com 153. FSE F5: Enter CAL mode by pressing PRINT. 4. F5.0 2: Use the TARE key to select 2 for cal in pounds and press PRINT to
ScaleResource.com 1617. StorE done: the meter now reboots to make use of the new constants. FIRST WEIGH DI-12/12K 1. Slide cal switch and push bu
ScaleResource.com 177. At this point, the accuracy of the cal can be checked by weight without leaving the cal mode. If the cal was accurate, pres
ScaleResource.com 186. Push enter 7. PLACE CAL WT, enter cal weight using keypad. 8. Place cal weight on scale and push enter. 9. Push clear/en
ScaleResource.com 192. PRESS ENTER- RESETS UPER DISPLAY TO ZERO AND ENABLES UNIT TO COUNT AT 10 TIMES NORMAL COUNT RATE. 3. RUN SUFFICIENT MATERIAL
ScaleResource.com 28. Remove the calibration weights and press the standby/operate key. 9. “……..” will appear for a couple of seconds, then CAL E
ScaleResource.com 20 THEM AT THIS TIME. 20. PRESS RUN. RESUME NORMAL OPERATION. MINOR CALIBRATION 1. ALLOW ½ HOUR WARM-UP WITH CONVEYOR BELT
ScaleResource.com 21 TRUCK(OR ACTUAL WEIGHT) ----------- X OLD = NEW RAMSEY(DISPLAYED WEIGHT) RICE LAKE IQ PLUS 310 1
ScaleResource.com 226. WVAL is displayed, edit the weight value to match your test weights then push enter. 7. WSPAN is displayed, place cal weig
ScaleResource.com 235. Place the known weight on the scale, if the reading is not equal to the cal weight, use the arrow keys to adjust reading. 6
ScaleResource.com 241. Push cal switch, display shows CONFIG 2. Using the SETPOINT key go to CALIBR. 3. Press the TIME/DATE key until WEIGHT is dis
ScaleResource.com 256. Use the three arrows to adjust the displayed value. Press ↑ to increase the flashing digit, ↓ to decrease it, → to m
ScaleResource.com 264 WORD LENGTH 7 BITS 8 BITS 5 PARITY CHECK YES NO 6 PARITY SE
ScaleResource.com 2711. 15 CAL 12. CAL d cal done 13. F1.3 0 expanded display, 0 is normal, 1 is expanded. 14. Push Print until SAVE is displ
ScaleResource.com 285. Press TEST E Scale. 6. Press TEST 7. ADD LOAD, push PRINT 8. 000000 9. TEST will move to location, TARE changes number.
ScaleResource.com 29 PrEC →→→→→→→→→→→→→→→→→→→→→→→→→→→→→→→→→→→→→→→→→→→→→→→→b.bal→→→→→→→→→→→→→→→→→→SPAn ↓ ↓↓↓↓
ScaleResource.com 3 A&D EK/ET 1. Make sure cal switch is on (not required for ET-120A) 2. Press MODE and RE-ZERO simultaneously. 3. CAL 0
ScaleResource.com 30 GROSS/NET ↓↓↓↓ SEALED/UNSEALED ↓↓↓↓ SET UP →→→→ADJUST→→→→END ↓ ZERO→→→→SPAN→→→→LINEAR→→→→DISPLAY→→→→END ↓
ScaleResource.com 413. Place the calibration on the platform and press the ZERO key, the display will show End, press the cal switch. A&D FC 1
ScaleResource.com 57. Place the cal weight on the platform that is specified by the display. These balances can only be calibrated using metric w
ScaleResource.com 64. Place 100g calibration mass on the center of the Trivet, 5. Wait for at least five seconds and press the TARE key. After a
ScaleResource.com 7 CARDINAL/DETECTO AP SERIES SN BEFORE 9302 1. Lift plastic display lens up and away from the front of the scale 2. Conn
ScaleResource.com 89. Place cal weight on scale and press Gross and ON/ZERO simultaneously. Display will show ------- then UNLD=. Remove the test
ScaleResource.com 95. Press print/enter to show current setting then using on/zero increment flashing digit to desired value and gross to switch t