A&D Digital Indicator AD-4531A Manuel d'utilisateur

Naviguer en ligne ou télécharger Manuel d'utilisateur pour Mesure, test et contrôle A&D Digital Indicator AD-4531A. AD-4531B Digital Indicator Simplified Instruction Manual [en] Manuel d'utilisatio

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 2
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 0
-2-
-
3
-
-
4
-
-1-
96
48
8.2
(138.4)
119.3
44.6
94.8
Min. 120
Min. 70
45
+0.6
-0
92
+0.8
-0
Panel cutout dimensions / Spacing
*Maintain the intervals above when installing
AD-4531B
Digital Indicator
Simplified Instruction Manual
© 2013 A&D Company, Limited. All rights reserved.
No part of this publication may be reproduced, transmitted,
transcribed, or translated into any language in any form by any
means without the written permission of A&D Company, Limited.
The contents of this manual and the specifications of the
instrument covered by this manual are subject to change for
improvement without notice.
This is a simplified instruction manual. For a detailed
instruction manual, visit the A&D website.
URL: http://www.aandd.jp/
1WMPD4002617
1. BEFORE USE
The AD-4531B digital indicator is a precision instrument. Unpack the
digital indicator carefully and confirm that all items are present.
1.1. Precautions Before Use
The AD-4531B is a precision instrument. Handle it with much care.
Avoid water and moisture.
Avoid vibration, shock, extremely high temperature and humidity,
direct sunlight, dust and air containing salt or sulfurous gases.
Avoid places where inflammable gases or vapors are present.
The operating temperature range is –10 °C to +40 °C.
A 100 VAC to 240 VAC power source is required. Use a stable
power supply free from sudden dropout or noise as they can cause
malfunctions. Avoid sharing the power line.
Keep cables away from power cables and other sources of
electrical noise.
Connect only a non-inductive load of 10 k or more to the analog
voltage output terminals.
Connect only a non-inductive load of 510 or less to the analog
current output terminals.
When connecting long cables to the sensors, keep the cables
away from power cables and other sources of electrical noise
Do not connect the AD-4531B to the power supply before
installation is completed. The AD-4531B has no switch to turn the
power off.
Use shielded load cell cables.
Do not connect too many sensors. Otherwise, instrument damage may occur.
1.2. Precautions During Use
The AD-4531B is a precision instrument that measures the microvolt
output from sensors. Prevent noise sources such as power lines,
radios, electric welders or motors from affecting the instrument.
Do not try to modify the AD-4531B.
In all hold modes, the hold data is saved digitally, so there is no
drooping of the value displayed on the display panel or the analog
output. Note that the hold function is disabled when the AD-4531B
is disconnected from the power supply.
Disconnect from the power supply before removing the cover.
When removing the cover, make sure that the power is
disconnected.
Do not touch the instrument immediately after it is disconnected
from the power supply.
To avoid electrical shock, do not touch the internal part of the instrument
within ten seconds after disconnecting from the power supply.
Be sure to fasten all the screws completely.
Loose screws may come off during operation and a short circuit
may occur or measurement errors may occur due to noise.
2. EXTERNAL DIMENSIONS
Unit: mm
3. GENERAL SPECIFICATIONS
Voltage requirement: 100 VAC to 240 VAC (50/60 Hz)
Power requirement: Approx. 10 VA
Sensor power supply 5 volt, 120 , one sensor maximum
5 volt, 350 , three sensors maximum
Operating temperature: -10 °C to +40 °C
Operating humidity: Max. 85% RH (no condensation)
External dimensions: 96 x 48 x 127.5 mm (W x H x D)
Panel cutout: 92 x 45 mm
Spacing: Width: 120 mm or more
Length: 70 mm or more
Weight: Approx. 290 g
4. OPTIONS
4.1. Options
AD-4530-200: Relay output
AD-4530-030: RS-485
AD-4530-040: RS-232C
AD-4530-007: Analog output
AD-4530-237: Relay output, RS-485, analog output
AD-4530-247: Relay output, RS-232C, analog output
Note: Only one option can be installed in the AD-4531B at a time.
4.2. Installing an Option
1. Remove the two screws that secure the guide rail, and then remove the
guide rail.
2. Remove the two screws that secure the case.
Pull the case out of the front panel (holding it as shown).
4. Remove unnecessary blank panels with nippers, etc.
5. Insert the option board at the position specified on the rear of the front
panel.
6. Reattach the case and guide rail by reversing the steps above.
5. FRONT PANEL
5.1. Display
Displays a measured or set value. To set the decimal point position, use
the function mode Cf-01. The display is composed of six seven-segment
LEDs plus a minus sign.
5.2. Status Indicators
Name Description
1 HI
Turns on when the measured value is greater than the
upper limit (HI).
2 OK
Turns on when the measured value is equal to or greater
than the lower limit and equal to or less than the upper
limit.
3 LO
Turns on when the measured value is less than the
lower limit (LO).
4 DZ Turns on when adjusting the zero value. (Digital zero)
5 HOLD Turns on when a value is being held.
5.3. Keys
Operation Function
Press to proceed to the upper limit value
setting mode.
When inputting a numerical value, press to
change the polarity.
Press to proceed to the lower limit value
setting mode.
When inputting a numerical value, press to
cancel the setting.
Press to turn the digital zero function on.
When inputting a numerical value, press to
shift the position of the blinking digit to the
right or change the function group.
+
Press to turn the digital zero function off.
Press to turn the hold function on or off.
When inputting a numerical value, press to
change the blinking digit or change the
function parameter.
Press to output the serial data (print).
When inputting a numerical value, press to
enter the setting.
+
Press to proceed to the calibration mode.
+
Press to proceed to the function selection
mode.
+ +
Press to proceed to the selection mode in
the check mode.
* The digital zero function sets a desired measurement point as zero and
displays the deviation from this zero point. When weighing a load, it can
be used as the tare, etc.
* The zero value is saved in non-volatile memory (EEPROM) and is
maintained even if the power is disconnected.
* To change the zero adjustment operation, use the function Cf-11.
To protect against accidental operation, press and to
turn the digital zero function on or off.
5.4. Operation Mode
- Upper / lower limit setting mode
Use this mode to set the upper and lower limit of the comparator.
- Calibration mode
Use this mode to perform zero and span calibration with an actual load
(calibration weight).
- Function mode
Use this mode to set functions.
- Check mode
Use this mode to confirm input and output operation.
6. REAR PANEL
Confirm the terminal numbers when making connections.
When making connections, confirm the terminal
numbers printed on the top of the indicator case.
6.1. Connector Function
6.1.1. AC input connector
(1) AC Connect the AC power cord.
(2) AC The power requirement is 100 VAC to 240 VAC, 50/60 Hz.
When making connections:
Switch off the power of all the instruments used.
Keep cables away from power cables and other
sources of electrical noise.
6.1.2. Load cell connector
(1) SHLD Connect the shielded cable of the sensor cable
(2) SIG- Negative signal input terminal for the sensors
(3) SIG+ Positive signal input terminal for the sensors
(4) EXC- Negative excitation terminal for the sensors
(5) SEN- Negative sensing input terminal for the sensors
(6) SEN+ Positive sensing input terminal for the sensors
(7) EXC+ Positive excitation terminal for the sensors
6.1.3. Control input/output connector
(1) EXT IN 1 Control input terminal 1
(2) EXT IN 2 Control input terminal 2
(3) IN COM Control input common terminal
(4) EXT OUT 1 Control output terminal 1
(5) EXT OUT 2 Control output terminal 2
(6) OUT COM Control output common terminal
6.1.4. Relay output connector (Option)
(1) COM Relay output common terminal
(2) LO Relay LO output terminal
Outputs LO when the measured value is less than the
lower limit.
(3) OK Relay OK output terminal
Outputs OK when the measured value is equal to or
greater than the lower limit and equal to or less than
the upper limit.
(4) HI Relay HI output terminal
Outputs HI when the measured value is greater
than the upper limit.
Comparator output
To prevent damage, do not exceed the rated capacities of
the output relays.
To protect the output relays, use a varistor, CR circuits or
diodes.
6.1.5. Serial communication connector (Option)
In the case of RS-232C
(1) IC Internally connected (Do not use)
(2) IC Internally connected (Do not use)
(3) SG Signal ground terminal
(4) TxD Sending terminal
(5) RxD Receiving terminal
In the case of RS-485 (Two-wire connection)
(1) B B terminal
(2) A A terminal
(3) SG Signal ground terminal
(4) B B terminal
(5) A A terminal
* Each of the A and B connections has two terminals.
These terminals are connected internally and can be used for a
terminating resistor or multi-drop connection.
6.1.6. Analog output connector (Option)
(1) COM Common terminal of the analog output
(2) IOUT Analog current output terminal
(3) VOUT Analog voltage output terminal
Front panel
Option board
Case
Guide rail
Blank panel
1 2 3 Display
4
5
Keys
4 3 21 54321 32 1
1 2 3 4
5
6 7 1 2 3 4 5 6 1 2
Relay output
RS
-
232
C
RS
-
485 Analog output
Load cell Control input
/
output
AC
Vue de la page 0
1 2

Résumé du contenu

Page 1

-2- -3--4--1- 9648 8.2 (138.4)119.344.694.8Min. 120 Min. 7045+0.6-0 92+0.8-0 Panel cutout dimensions / Spacing*Maintain the intervals

Page 2 - 8. FUNCTION MODE

-6- -7--8--5- 7. CALIBRATION The AD-4531B measures voltage signals from sensors and displays the values. Calibration is performed so that the AD-453

Commentaires sur ces manuels

Pas de commentaire